Moisture Analyser MA100
-Infrared heating using a round ceramic heating element (MA100C), a halogen round heater (MA100H) or a coiled quartz (CQR) heating element (MA100Q)
-Utmost safety due to self-calibrating system and full GMP suitability
-Ideally suited for samples with high moisture content due to quick control behaviour
-Sample quantity 5-15g
-% moisture content (optionally available with conversion factor), % dry weight, RATIO value, mg weight loss, g residue, g/kg residue, g/l residue
-Temperature range and settings 30°C to 180°C, stand-by temperature selectable from 30°C to 100°C in increments of 1 degree
-Heating programs: standard, quick, gentle and phase drying
-Shutoff criteria (optional): fully automatic, SPRM mode (optimization to a reference system), semi-automatic (1-50mg/5-300sec.), semi-automatic (0.1-5.0%/5-300sec.), timer setting (0.1-999min.), timer setting and fully/semi-automatic (0.1-999min. and automatic), manual
-Access to sample chamber Motorized cover
-30 programs saved in a non-volatile memory (freely selectable parameters)
-Result statistics for the last 9999 measurements/program
-Menu-driven, alphanumeric dialogue text (5 languages selectable)
-Optimized text entry via softkeys, numeric entry via keypad, optional data entry via barcode scanner, printout optional using an internal printer (Cat. No. 9.902 611), configurable GLP-report for measuring results, calibration, adjustment (5 languages selectable)
-Internal calibration weight, automatic reproTEST for the weighing system, automatic temperature calibration with optional temperature adjustment set (Cat. No. 6.264 042) , documentation in accordance with the applicable quality assurance requirements
-Safety Parameter settings password-protected against unauthorized access
Specifications
Readability: 0.01 % or 0.001 g / 0.001 % or 0.0001 g
Weighing range, max: 100 g
Reproducibility, weight of sample 1g:
Reproducibility, weight of sample 5g: ±0,1 %
±0,02 %
Temperature range: 30 ... 180 °C, in increments of 1 °C
Power consumption: 700 VA
Dimensions (W x D x H): 350 x 453 x 156 mm
Weight: approx. 8 kg
Power supply: 230 V, 50 ... 60 Hz
Tuotekoodi | Valmistaja | Nimi | Type | max. load (g) | Resolution (g) | Min. temp. (°C) | Max. temp. (°C) | Width (mm) | Depth (mm) | Height (mm) | Heating technology | Repeatability at 1 g (%) | Repeatability at 5 g (%) |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
9902602 | MA100C-000230V1 Sartorius Lab Instruments |
Humidity analyzer MA100C, 100 g/0.1 g 1 |
MA100C | 100 | 0.001000 | 30 | 180 | 350 | 453 | 156 | Ceramic radiator | 0.10 | 0.02 |
9902603 | MA100H-000230V1 Sartorius Lab Instruments |
Humidity analyzer MA100H 1 |
MA100H | 100 | 0.001000 | 30 | 180 | 350 | 453 | 156 | Halogen radiator | 0.10 | 0.02 |
9902604 | MA100Q-000230V1 Sartorius Lab Instruments |
Humidity analyzer MA100Q 1 |
MA100Q | 100 | 0.001000 | 30 | 230 | 350 | 453 | 156 | Quartz glass radiator | 0.10 | 0.02 |